HPLC Protocols - including fluorescence detection
Supplies Companies
Column Care
HPLC systems
Alltech (or
here
for non-Java version, probably less prone to crashing): For pricing
info (Online Store), use account
CA095.
But don't order online.
Vydac
Storage conditions: >50% organic in H2O (NOT BUFFER). Absolutely no TFA.
Exposure of column to pH6.5 or greater: Use silica saturation column between pump and injector, to minimize loss of silica from column (and resulting void volume and decrease in resolution).
Agilent Technologies (formerly Hewlett-Packard)
The PE 650-10 S is dead.
Make sure the flow cell flows from bottom right to top left, when
looking from front to back of instrument along excitation beam path. If
the flow is from bottom left to top right, flip the flow cell upside-down.
I know, it doesn't make sense, but otherwise you will probably get
bubbles trapped in the flow cell. If the flow is from top to bottom:
get with the program, you don't know what you're doing!
Initial suggested slitwidths are 10 nm.
You need to make sure the Instrument/Setup ADC Signals/Stop Datatime
setting is the same as the end of the run time.